TY - JOUR T1 - Accuracy of blood-pressure monitors owned by patients with hypertension (ACCU-RATE study): a cross-sectional, observational study in central England JF - British Journal of General Practice JO - Br J Gen Pract DO - 10.3399/bjgp20X710381 SP - bjgp20X710381 AU - James A Hodgkinson AU - Mei-Man Lee AU - Siobhan Milner AU - Peter Bradburn AU - Richard Stevens AU - FD Richard Hobbs AU - Constantinos Koshiaris AU - Sabrina Grant AU - Jonathan Mant AU - Richard J McManus Y1 - 2020/06/02 UR - http://bjgp.org/content/early/2020/06/01/bjgp20X710381.abstract N2 - Background Home blood-pressure (BP) monitoring is recommended in guidelines and is increasingly popular with patients and health professionals, but the accuracy of patients’ own monitors in real-world use is not known.Aim To assess the accuracy of home BP monitors used by people with hypertension, and to investigate factors affecting accuracy.Design and setting Cross-sectional, observational study in urban and suburban settings in central England.Method Patients (n = 6891) on the hypertension register at seven practices in the West Midlands, England, were surveyed to ascertain whether they owned a BP monitor and wanted it tested. Monitor accuracy was compared with a calibrated reference device at 50 mmHg intervals between 0–280/300 mmHg (static pressure test); a difference from the reference monitor of +/−3 mmHg at any interval was considered a failure. Cuff performance was also assessed. Results were analysed by frequency of use, length of time in service, make and model, monitor validation status, purchase price, and any previous testing.Results In total, 251 (76%, 95% confidence interval [95% CI] = 71 to 80%) of 331 tested devices passed all tests (monitors and cuffs), and 86% (CI] = 82 to 90%) passed the static pressure test; deficiencies were, primarily, because of monitors overestimating BP. A total of 40% of testable monitors were not validated. The pass rate on the static pressure test was greater in validated monitors (96%, 95% CI = 94 to 98%) versus unvalidated monitors (64%, 95% CI = 58 to 69%), those retailing for >£10 (90%, 95% CI = 86 to 94%), those retailing for ≤£10 (66%, 95% CI = 51 to 80%), those in use for ≤4 years (95%, 95% CI = 91 to 98%), and those in use for >4 years (74%, 95% CI = 67 to 82%). All in all, 12% of cuffs failed.Conclusion Patients’ own BP monitor failure rate was similar to that demonstrated in studies performed in professional settings, although cuff failure was more frequent. Clinicians can be confident of the accuracy of patients’ own BP monitors if the devices are validated and ≤4 years old. ER -