Mercury devices (13% of all devices identified) | 79 | 75 | | 5% (4/75 amber, 0/75 red) |
Accoson | 74 (94) | 70 | Four: one with bubble in mercury, stored in drawer and unclaimed, last tested 1997, presumed not in use; three labelled as out of use since previous test, but wall-mounting and cost of decommissioning precluded removal from premises | 6% (4/70 amber, 0/70 red) |
Other manufacturers | 5(6) | 5 | | 0% (0/5 amber/red) |
|
Aneroid devices (32% of all devices identified) | 192 | 191 | | 22% (37/191 amber, 4/191 red) |
Accoson | 69 (36) | 69 | | 14% (10/69 amber, 0/69 red) |
Welch-Allyn | 46 (24) | 46 | | 15% (7/46 amber, 0/46 red) |
boso | 13(7) | 13 | | 8% (1/13 amber, 1/13 red) |
Other device manufacturer | 34(18) | 33 | One with missing part (Riester Big Ben) | 18% (6/33 amber, 0/33 red) |
Devices with pharmaceutical branding | 23(12) | 23 | | 52% (11/23 amber, 2/23 red) |
Unknown manufacturer | 7(4) | 7 | | 43% (2/7 amber, 1/7 red) |
|
Digital devices (53% all devices identified) | 323 | 308 | 15 | 12% (36/308 amber, 0/308 red) |
Omron | 198(61) | 196 | Two wrist monitors — incompatible with testers | 12% (23/196 amber, 0/196 red) |
A&D/bosoa | 104(32) | 103 | One UB511 wrist monitor — incompatible with testers | 11%(11/103 amber, 0/103 red) |
Other manufacturer | 11(3) | 6 | Five: either incompatible with testers (3) — H21188U Wrist, KD 525, Rossmax; or mechanically failed (2) — Transtec, Prestige | 17% (1/6 amber, 0/6 red) |
High street pharmacistb | 10(3) | 3 | Seven: either incompatible with testers (6), or mechanically failed (1), all Lloyds Pharmacy | 33% (1/3 amber, 0/3 red) |
|
Hybrid devices (2% of all devices identified) | 10 | 10 | | |
A&Dc | 10(100) | 10 | | 0% (0/10 amber/red) |