The influence of experience on the reproducibility of the ankle-brachial systolic pressure ratio in peripheral arterial occlusive disease

Eur J Vasc Endovasc Surg. 1999 Jul;18(1):25-9. doi: 10.1053/ejvs.1999.0843.

Abstract

Objectives: to estimate the intra-observer variability of the measurement of the ankle-brachial systolic pressure index (ABPI) and to compare the reproducibility of the measurements by experienced vascular laboratory assistants and by less-experienced general practice personnel.

Design: repeated measurement of ABPI by general practitioners (GPs), GP-assistants and vascular laboratory assistants using a pocket Doppler device and a random-zero sphygmomanometer.

Methods and materials: ABPI was measured in six patients with various degrees of PAOD by two experienced observers (vascular laboratory assistants) and by 24 less-experienced observers (18 practice assistants, six GPs).

Results: the total number of measurements was 354. The overall intra-observer variability estimate was 11.8% ABPI. The intra-observer variability was 7.3% in the experienced observers and 12.0% in the less-experienced observers. The difference of variability between experienced and less-experienced observers was significant.

Conclusions: the ABPI is suitable in follow-up studies where repeated measurements are needed. Differences between measurements can be minimised by performing repeated measurements or by using more experienced observers.

MeSH terms

  • Aged
  • Ankle / blood supply*
  • Arterial Occlusive Diseases / physiopathology*
  • Blood Pressure*
  • Brachial Artery / physiopathology*
  • Clinical Competence*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Observer Variation
  • Reproducibility of Results
  • Sphygmomanometers
  • Statistics as Topic