Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use?

Blood Press Monit. 2005 Feb;10(1):51-6. doi: 10.1097/00126097-200502000-00009.

Abstract

Objective: To assess the measurement consistency of SpaceLabs ambulatory recorders (Spacelabs, Washington, USA) that are in regular use.

Methods: A total of 14 SpaceLabs 90207 and one 90217 ambulatory recorders were tested for measurement consistency using the Dynatech CuffLink (Dynatech, Nevada, USA), a commercially available non-invasive blood pressure (NIBP) simulator. The NIBP recorders were tested at a range of pressures with 20 repeated determinations at a simulated 120/80 mmHg and five repeated determinations at simulated pressures of 80/50, 100/80, 150/100, 200/165 and 250/195 mmHg. Tests were carried out in 1998, 2002 and late 2003 or early 2004.

Conclusions: All 15 SpaceLabs recorders measured consistently over the 6 years with 89.5% of the differences in average pressures, recorded by any particular device at each recorded pressure, less than 2 mmHg between successive test episodes. The maximum difference was 4.5 mmHg and 60.1% of the differences were less than 1 mmHg. The measurements for all devices were within the tolerances specified by the supplier for the device when tested with the simulator. Maintenance records also show that most devices required breakdown maintenance less than once every 3 years. The results show that the SpaceLabs devices maintain measurement consistency in the demanding conditions of ambulatory pressure recording over several years.

MeSH terms

  • Aerospace Medicine / instrumentation*
  • Aerospace Medicine / standards
  • Blood Pressure
  • Blood Pressure Monitoring, Ambulatory / standards
  • Blood Pressure Monitors / standards*
  • Equipment Failure
  • Humans
  • Reproducibility of Results
  • Time Factors